Shandong Fengtu IOT Technology Co., Ltd
Sales Manager:Ms. Emily Wang
Cel,Whatsapp,Wechat:+86 15898932201
Email:info@fengtutec.com
Add:No. 155 Optoelectronic Industry Accelerator, Gaoxin District, Weifang, Shandong, China
Sales Manager:Ms. Emily Wang
Cel,Whatsapp,Wechat:+86 15898932201
Email:info@fengtutec.com
Add:No. 155 Optoelectronic Industry Accelerator, Gaoxin District, Weifang, Shandong, China
Model:FT-CW16H
Brand:fengtu
Basic structure of fluorescence quantitative PCR instrument
The fluorescence quantitative PCR instrument is an instrument for real-time detection of reactions. It mainly consists of a gene amplification thermal cycle system, a fluorescence real-time detection system, a microcircuit control system, a computer and application software. Two of the core functional modules are: thermal circulation system and fluorescence real-time detection system. The working principle of the gene amplification thermal cycle system is basically the same as that of the traditional gene amplification instrument. It uses semiconductor heating and cooling to complete the thermal cycle process. The fluorescence detection system mainly consists of fluorescence excitation components, optical signal transmission components, fluorescence detection components, and control systems.
fluorescence quantitative PCR instrument technical specifications
The "///" symbol indicates that the national standard currently does not define this indicator or mandatory requirement.
Indicator name | CW16H | CW32H | CW48H | |
Appearance | ||||
Overall dimensions | 235mm*385mm*175mm (width*depth*height) | |||
weight | 5.6kg | 5.7kg | 5.8Kg | |
Electrical parameters | ~220V/50Hz, 255W | |||
Data interface | USB 2.0 *2 (two on the right) | |||
Environmental parameters | ||||
Operating conditions | Temperature: 10-30℃, humidity: 20%~80% | |||
Shipping and storage conditions | Temperature: -20~55℃, humidity: 20%~80% | |||
altitude | <2500 meters | |||
Noise level | A-weighted, <60dB | |||
Sample parameters | ||||
sample size | 16*0.2mL | 32*0.2mL | 48*0.2mL | |
Test tube type | Single tube, eight row test tubes | |||
sample volume | 15-100uL | |||
Temperature characteristics | ||||
Heating/cooling method | Semiconductor heating/cooling | |||
temperature range | 4℃-99℃ | |||
Maximum heating rate | ≥3.5℃/s | |||
average heating rate | ≥2.5℃/s | |||
Maximum cooling rate | ≥3.5℃/s | |||
average cooling rate | ≥2.5℃/s | |||
Temperature control accuracy | ≤±0.01℃ | |||
Temperature accuracy | ≤±0.1℃ | |||
temperature uniformity | ≤±0.3℃ | |||
Optical properties | ||||
Number of fluorescence detection channels | 4 channels | |||
Light emitting device | High brightness LED | |||
lighting devices | High sensitivity, high signal-to-noise ratio photodiode | |||
Adapter probe or dye | First channel: 470/520 FAM, SYBR Green | |||
Second channel: 530/570 HEX, JOE, VIC | ||||
Third channel: 580/610 ROX, CY3.5, Texas-Red | ||||
Fourth channel: 630/670 CY5 | ||||
Detection sensitivity | 1 copy | |||
Linear detection range | 100~1010 copies | |||
linear correlation coefficient | ≥0.999 | |||
Channel crosstalk | No crosstalk | |||
Test repeatability | ≤1.0% |
Main application fields of fluorescence quantitative PCR instrument
Foodborne pathogen testing
food safety
Animal and plant origin testing
Plant Science and Agricultural Biotechnology
Gene mutations and polymorphisms
animal diseases
basic scientific research
water body monitoring
Pathogen nucleic acid detection
GMO testing
Animal epidemic surveillance
Food hygiene and quarantine
Customs import and export quarantine
In modern society, the measurement of visibility is of great significance. In terms of meteorological research, visibility data is a crucial element for analyzing atmospheric characteristics. Its value can reflect the conditions of water vapor and particulate matter in the atmosphere, helping meteor...
How to measure "visibility"?I. Visual Estimation MethodThis is a rather traditional and simple approach. Observers use their naked eyes to observe target objects. For example, during the day, they select mountains, buildings, etc. that have a sharp contrast in color and shape with the back...
It is feasible to use soil moisture monitoring system to guide agricultural irrigation. Soil moisture monitoring systems can monitor soil moisture, temperature, conductivity and other parameters in real time to help farmers understand the soil condition of farmland in time and provide scientific bas...
Fengtu Technology has 4 water level monitoring stations, the models are: FT-SW1, FT-SW2, FT-SW3, FT-SW4. For details, please contact the customer service of this website.Fengtu Hydrological Monitoring Station adopts radar technology to measure hydrological data by non-contact method, with low power...